According to standard IEC 62058-31

Paragraph Preliminary tests and pre-conditioning


The High accuracy static meters need to be warmed up before accuracy testing according to acceptance testing international standard requirements. 15 - 20 min is the minimum time for this step.

We hold each meter at least 20 min at warm up before final testing and calibration procedure.